EasyManua.ls Logo

Agilent Technologies G1314B - Flow Connections to the Detector

Agilent Technologies G1314B
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46 Agilent 1260 Infinity VWD User Manual
3 Installing the Detector
Flow Connections to the Detector
Flow Connections to the Detector
Tools required Description
Wrench, 1/4 – 5/16 inch
(for capillary connections)
Parts required # p/n Description
1 G1314-68755 Accessory kit
Hardware required Other modules depend on system setup
Preparations Detector is installed in the LC system.
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
The handling of solvents, samples and reagents can hold health and safety risks.
When working with these substances observe appropriate safety procedures (for
example by wearing goggles, safety gloves and protective clothing) as described in
the material handling and safety data sheet supplied by the vendor, and follow good
laboratory practice.
The volume of substances should be reduced to the minimum required for the
analysis.
Do not operate the instrument in an explosive atmosphere.
NOTE
The flow cell is shipped with a filling of isopropanol (also recommended when the
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.

Table of Contents

Related product manuals