Agilent 1260 Infinity VWD User Manual 89
Error Information
7
General Error Messages
Lost CAN Partner
Error ID: 0071
During an analysis, the internal synchronization or communication
between one or more of the modules in the system has failed.
The system processors continually monitor the system configuration. If one
or more of the modules is no longer recognized as being connected to the
system, the error message is generated.
Probable cause Suggested actions
1 Not-ready condition in one of the
instruments connected to the remote line.
Ensure the instrument showing the not-ready
condition is installed correctly, and is set up
correctly for analysis.
2 Defective remote cable.
Exchange the remote cable.
3 Defective components in the instrument
showing the not-ready condition.
Check the instrument for defects (refer to the
instrument’s documentation).
Probable cause Suggested actions
1 CAN cable disconnected.
Ensure all the CAN cables are connected
correctly.
Ensure all CAN cables are installed
correctly.
2 Defective CAN cable.
Exchange the CAN cable.
3 Defective main board in another module.
Switch off the system. Restart the system, and
determine which module or modules are not
recognized by the system.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B and is the answer not in the manual?

Agilent Technologies G1314B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B
CategorySecurity Sensors
LanguageEnglish

Related product manuals