Agilent 1260 Infinity VWD User Manual 113
Test Functions
8
Dark Current Test
Dark Current Test Failed
Probable cause Suggested actions
1 Defective sample or reference diode.
Please contact your Agilent service
representative.
2 Defective sample or reference ADC board.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.