92 Agilent 1260 Infinity VWD User Manual
7 Error Information
General Error Messages
Compensation Sensor Short
Error ID: 0080
The ambient- compensation sensor (NTC) on the main board in the module
has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
main board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient
temperature changes. If the resistance across the sensor falls below the
lower limit, the error message is generated.
Fan Failed
Error ID: 0068
The cooling fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the
fan speed. If the fan speed falls below a certain limit for a certain length
of time, the error message is generated.
This limit is given by 2 revolutions/second for longer than 5 seconds.
Depending on the module, assemblies (e.g. the lamp in the detector) are
turned off to assure that the module does not overheat inside.
Probable cause Suggested actions
1 Defective main board.
Please contact your Agilent service
representative.