20 Agilent 1260 Infinity VWD User Manual
2 Site Requirements and Specifications
Site Requirements
Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the
instrument.
Power Consideration
The detector power supply has wide ranging capabilities, see “Physical
Specifications” on page 24. It accepts any line voltage in the above
mentioned range. Consequently, there is no voltage selector in the rear of
the detector. There are also no externally accessible fuses, because
automatic electronic fuses are implemented in the power supply.
Instrument is partially energized when switched off
The power supply still uses some power, even when the power switch on the front
panel is turned OFF. Repair work at the detector can lead to personal injuries, e. g.
shock hazard, when the detector cover is opened and the instrument is connected to
power.
➔ To disconnect the detector from the power line, unplug the power cord.
Hazard of electrical shock or damage of your instrumentation
can result, if the devices are connected to a line voltage higher than specified.
➔ Connect your instrument to the specified line voltage only.