134 Agilent 1260 Infinity VWD User Manual
9 Maintenance
Replacing the Interface Board
Replacing the Interface Board
When When defective or for installation of the board or for all repairs inside the detector.
Parts required # p/n Description
1 G1351-68701 Interface board (BCD) with external contacts and BCD outputs
OR 1 G1369B or
G1369-60002
Interface board (LAN)
OR 1 G1369C or
G1369-60012
Interface board (LAN)
1 Install the ESD strap. Move the power lock across the
power inlet.
2 If required, unscrew and remove the interface board.
Place the board on the ESD kit.
Next Steps:
3 If required, insert the interface board and fix the screws.
4 Remove the ESD strap.
5 Reinstall the module into the stack.
:H9higVe
EdlZgadX`