96 Agilent 1260 Infinity VWD User Manual
7 Error Information
Detector Error Messages
No heater current
Error ID: 7453
The lamp heater current in the detector is missing. During lamp ignition,
the processor monitors the heater current. If the current does not rise
above the lower limit within 1 , the error message is generated.
Probable cause Suggested actions
1 Lamp disconnected.
Ensure the lamp is connected.
2 Ignition started without the top foam in
place.
Please contact your Agilent service
representative.
3 Fan not running (permitting lamp on).
Please contact your Agilent service
representative.
4 Defective main board.
Please contact your Agilent service
representative.
5 Defective or non-Agilent lamp.
Exchange the lamp.
6 Defective power supply.
Please contact your Agilent service
representative.

Table of Contents