Agilent 1260 Infinity VWD User Manual 97
Error Information
7
Detector Error Messages
Wavelength calibration setting failed
Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed: Zero-order calibration failed.
Calibration 1 Failed: 656 nm calibration failed.
Probable cause Suggested actions
1 Lamp is OFF.
Switch on the lamp.
2 Incorrect flow cell installation.
Ensure the flow cell is installed correctly.
3 Flow cell contamination or air bubbles.
Clean/replace flow cell windows or remove air
bubbles.
4 Intensity too low.
Replace lamp.
5 Current step value too far from maximum.
Repeat the calibration.
Please contact your Agilent service
representative.
6 Misaligned/defective grating assembly.
Please contact your Agilent service
representative.
7 Defective main board.
Please contact your Agilent service
representative.