Agilent 1260 Infinity VWD User Manual 47
Installing the Detector
3
Flow Connections to the Detector
1
Press the release buttons and remove the front cover to
have access to the lamp area.
2 Remove the plastic dummy cover.
OR
If the detector uses a metal plate, loose the screws of the
flow cell dummy plate by turning each screw one turn.
Then unscrew the screws completely. This is required to
avoid any problem with helicoil insert in the casting.
3 Press the flow cell completely into the slot and tighten
the cell screws (both parallel) until the mechanical stop.
4 Assemble the column-detector capillary. Depending on
the flow cell type it may be a PEEK or SST capillary.