EasyManua.ls Logo

Agilent Technologies G1314B - Page 131

Agilent Technologies G1314B
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent 1260 Infinity VWD User Manual 131
Maintenance
9
Using the Cuvette Holder
1
Locate the cuvette holder on the desk. 2 Unscrew the bracket.
3 Insert the cuvette with the sample into the holder. The
clear side of the cuvette must be visible.
4 Replace the bracket and fix the cuvette.
Next Steps:
5 Install the cuvette holder in the instrument.
6 Perform your Wavelength Verification/Calibration (see “Wavelength Verification-Calibration” on page 108) to check the
correct position of the cuvette holder.
A^\]ieVi]
8aZVgh^YZ

Table of Contents

Related product manuals