Agilent 1260 Infinity VWD User Manual 23
Site Requirements and Specifications
2
Site Requirements
Environment
Your detector will work within the specifications at ambient temperatures
and relative humidity described in “Physical Specifications” on page 24.
ASTM drift tests require a temperature change below 2 °C/h (3.6 °F/h)
over one hour period. Our published drift specification (refer also to
“Performance Specifications G1314B” on page 25 or “Performance
Specifications G1314C” on page 27) is based on these conditions. Larger
ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency
and the amplitude of the temperature changes to below 1 °C/h (1.8 °F/h).
Turbulences around one minute or less can be ignored.
Condensation within the module
Condensation will damage the system electronics.
➔ Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔ If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
This module is designed to operate in a typical electromagnetic environment, i.e. where RF
transmitters such as mobile telephones may not be used in close proximity.