22 Agilent 1260 Infinity VWD User Manual
2 Site Requirements and Specifications
Site Requirements
Bench Space
The detector dimensions and weight (see “Physical Specifications” on
page 24) allows you to place the detector on almost any desk or
laboratory bench. It needs an additional 2.5 cm (1.0 inch) of space on
either side and approximately 8 cm (3.1 inch) in the rear for air
circulation and electric connections.
If the bench should carry an Agilent 1200 Infinity Series system, make
sure that the bench is designed to bear the weight of all modules.
The detector should be operated in a horizontal position.
Unintended use of supplied power cords
Using power cords for unintended purposes can lead to personal injury or damage of
electronic equipment.
➔ Never use the power cords that Agilent Technologies supplies with this instrument
for any other equipment.