Agilent 1260 Infinity VWD User Manual 27
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications G1314C
Tabl e 4 Performance Specifications G1314C
Type Specification Comments
Detection type Double-beam photometer
Light source Deuterium lamp
Wavelength range 190 – 600 nm
Short term noise (ASTM)
0.5·10
-5
AU at 254 nm
See “Specification
Conditions” on page 29
Drift
3·10
-4
AU/h at 254 nm
See “Specification
Conditions” on page 29.
Linearity > 2 AU (5 %) upper limit See “Specification
Conditions” on page 29.
Wavelength accuracy ± 1 nm Self-calibration with
deuterium lines, verification
with holmium oxide filter
Maximum data rate 55 Hz
Band width 6.5 nm typical
Flow cells Standard: 14 µL volume,
10 mm cell path length and
40 bar (580 psi) pressure
maximum
High pressure: 14 µL volume,
10 mm cell path length and
400 bar (5800 psi) pressure
maximum
Micro: 1 µL volume, 5 mm cell
path length and 40 bar
(580 psi) pressure maximum
Semi-micro: 5 µL volume,
6 mm cell path length and
40 bar (580 psi) pressure
maximum
Can be repaired on component
level
Control and data evaluation Agilent ChemStation for LC