26 Agilent 1260 Infinity VWD User Manual
2 Site Requirements and Specifications
Performance Specifications
Control and data evaluation Agilent ChemStation for LC
Analog outputs Recorder/integrator: 100 mV
or 1 V, output range 0.001 to
2 AU, one output
Communications Controller-area network
(CAN), RS-232C, APG Remote:
ready, start, stop and
shut-down signals, LAN
(optional)
Safety and maintenance Extensive diagnostics, error
detection and display (through
Agilent ChemStation), leak
detection, safe leak handling,
leak output signal for
shutdown of pumping system.
Low voltages in major
maintenance areas.
GLP features Early maintenance feedback
(EMF) for continuous tracking
of instrument usage in terms
of lamp burn time with
user-settable limits and
feedback messages.
Electronic records of
maintenance and errors.
Verification of wavelength
accuracy with built-in holmium
oxide filter.
Housing All materials recyclable.
Tabl e 3 Performance Specifications G1314B
Type Specification Comments

Table of Contents