186 Agilent 1260 Infinity VWD User Manual
12 Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode
with default parameter settings.
If you use the following switch settings and power the instrument up
again, a forced cold start has been completed.
CAUTION
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
Save your methods and data before executing a forced cold start.
Tabl e 26 Forced Cold Start Settings (without on-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
TEST/BOOT11000001