Agilent 1260 Infinity VWD User Manual 41
Installing the Detector
3
Installation Information on Leak and Waste Handling
1 Stack the modules according to the adequate stack configuration.
The leak pan outlet of the upper module must be vertically positioned
above the leak tray of the lower module, see Figure 9 on page 40.
2 Connect data and power cables to the modules, see section Installing
the Module below.
3 Connect capillaries and tubes to the modules, see section Flow
Connections to the module below or the relevant system manual.
1 Solvent cabinet
2Leak pan
3 Leak pan's outlet port (A), leak funnel (B) and corrugated waste tube (C)
4 Waste tube of the sampler’s needle wash
5 Condense drain outlet of the autosampler cooler
6 Waste tube of the purge valve
7 Waste tube
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
Keep solvent path free from blockages.
Keep the flow path closed (in case the pump in the system is equipped with a
passive inlet valve, solvent may leak out due to hydrostatic pressure, even if your
instrument is off).
Avoid loops.
Tubes must not sag.
Do not bend tubes.
Do not immerse tube end in waste liquid.
Do not intubate tubes in other tubes.
For correct tubing follow instructions on label attached to the module.