72 Agilent 1260 Infinity VWD User Manual
4 Using the Detector
Special Settings of the Detector
Tabl e 9 Peak Width — Response Time — Data Rate (G1314B VWD)
Peak Width (min)
1
1
Values in the User Interface may be rounded.
Response Time (s) Data Rate (Hz)
<0.005 0.12 13.74
>0.005 0.12 13.74
>0.01 0.25 13.74
>0.025 0.5 13.74
>0.05 1.0 6.87
>0.10 2.0 3.43
>0.20 4.0 1.72
>0.40 8.0 0.86
Tabl e 10 Peak Width — Response Time — Data Rate (G1314C VWD SL)
Peak Width (min)
1
1
Values in the User Interface may be rounded.
Response Time (s) Data Rate (Hz)
<0.00125 <0.031 55
>0.00125 0.031 27.5
>0.0025 0.062 13.74
>0.005 0.12 13.74
>0.01 0.25 13.74
>0.025 0.5 13.74
>0.05 1.0 6.87
>0.10 2.0 3.43
>0.20 4.0 1.72
>0.40 8.0 0.86