1260 Infinity II Flexible Pump User Manual 81
3Using the Module
Setting up the Pump with the Instrument Control Interface
Timetable Settings
Use the Timetable to program changes in the pump parameters during the
analysis by entering a time in the Time field and appropriate values in the
following fields of the timetable. Changes in flow rate occur linearly from either
time zero or the time of the last defined change; other parameters change
instantaneously at the time defined in the timetable.
Show Advanced Timetable toggles the timetable display between standard mode
and advanced mode.
The following parameters can be changed:
•Change Contacts
•Change Flow
• Change Max. Pressure Limit
• Change Solvent Composition - You can only use solvents, which have been
enabled in the solvents section.
• Function centric view - This check box allows you to display parameter
changes instead of a time table.