Agilent InfinityLab LC Series Diode Array Detectors User Manual 101
5 Optimizing the Detector
Optimizing the Detector Regarding to the System
> 0.05 1 5 5 5 5 5
> 0.1 2 2.5 2.5 2.5 2.5 2.5
> 0.2 4 1.25 1.25 1.25 1.25 1.25
> 0.4 8 0.625 0.625 0.625 0.625 0.625
> 0.85 16 0.3125 0.3125 0.3125 0.3125 0.3125
1
Values in the User Interface may be rounded.
Table 20 Peak Width — Response Time — Data Rate (G7117A/G7117C)
Peak width at half
height [min]
1
1
Values in the User Interface may be rounded.
Response [s] Scan data
rate[Hz] ≤251
pts/scan
Scan data
rate[Hz] ≤501
pts/scan
Scan data
rate[Hz] > 501
pts/scan
< 0.0015625 0.015625 120 120 40 20
> 0.0015625 0.03125 120 120 40 20
> 0.003125 0.0625 80 80 40 20
> 0.00625 0.125 40 40 40 20
> 0.0125 0.25 20 20 20 20
> 0.025 0.5 10 10 10 10
> 0.05 1 5 5 5 5
> 0.1 2 2.5 2.5 2.5 2.5
> 0.2 4 1.25 1.25 1.25 1.25
> 0.4 8 0.625 0.625 0.625 0.625
> 0.85 16 0.3125 0.3125 0.3125 0.3125
Table 19 Peak Width — Response Time — Data Rate (G7117B)
Peak width at half
height [min]
1
Response
[s]
Signal data
rate [Hz]
Scan data rate
[HZ]
≤126 pts/scan
Scan data rate
[HZ]
≤251 pts/scan
Scan data rate
[HZ]
≤501 pts/scan
Scan data rate
[HZ]
>501 pts/scan
NOTE
The maximum spectra scan rate depends on the data points per scan, see
Table 19 on page 100 and Table 20 on page 101. Running at 240, the spectra
scan data rate is reduced automatically if the spectra scan data rate is more than
251 points/scan.