Agilent InfinityLab LC Series Diode Array Detectors User Manual 73
4 Preparing the Module
Installing Capillaries
For correct installation of capillary connections it's important to choose the
correct fittings, see “Syntax for Capillary Description” on page 68.
Table 14 Capillary connections for 1290 Infinity II Bio LC
p/n From To
Bottle Head Assembly (G7120-60007) Solvent Bottle Infinity II Pump
Capillary MP35N 0.17 mm x 500 mm, SI/SI (5500-1419) Pump Multisampler
Capillary MP35N 0.12 mm x 500 mm SI/SI (5500-1279) Multisampler MCT
Quick Connect Capillary MP35N 0.12 mm x 105 mm
(5500-1578)
MCT Heat Exchanger Column
Quick Turn Capillary MP35N 0.12 mm x 280 mm (5500-1596) Column/MCT Valve Detector
Waste accessory kit (5062-8535) VWD Waste
Tube PTFE 0.7 mm x 5 m, 1.6 mm od (5062-2462) DAD/FLD Waste
Analytical tubing kit 0.25 mm i.d. PTFE-ESD (G5664-68712) Detector Fraction Collector
1 Select a nut that is long enough for the fitting you'll be
using.
2 Slide the nut over the end of the tubing or capillary.