60 Agilent Intuvo 9000 GC Installation
1 Installing the GC
REMOTE connector
Provides a port to remotely start and stop other instruments
using the APG protocol. A maximum of 10 instruments can be
synchronized using this connector. See “Using the Remote
Start/Stop Cable” on page 100 for more detail.
EVENT connector
This connector provides two passive contact closures and two
24-volt outputs for controlling external devices.
BCD input connector
This connector provides two control relays and a BCD input for
a stream selection valve or a BCD generating device.
LAN connector
Standard Local Area Network (LAN) connector, for
communication with data systems and other devices via TCP/IP.
CAUTION
This connector is similar to the EVENT connector. Plugging a
non-BCD cable into the BCD connector can damage the GC.