EasyManua.ls Logo

Agilent Technologies U1252B - Page 151

Agilent Technologies U1252B
191 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests and Calibration 6
Agilent U1251B/U1252B User’s and Service Guide 131
Step Test Function Range 5520A Output
Error from nominal 1 year
U1251B U1252B
1000.0 μF 1000.0 μ 10.5 μ 10.5 μF
10.00 mF 10.00 mF ± 0.105 mF ± 0.105 mF
100.00 mF 10.00 mF ± 0.4 mF ± 0.4 mF
14 Press button to go to
mode
[8][13]
–200 ºC until
1372 ºC
0 ºC ± 3 ºC ± 3 ºC
100 ºC ± 3.3 ºC ± 3.3 ºC
15 Turn the rotary swich to the
position
500 μA500 μA ± 0.55 μA
[9]
± 0.3 μA
[9]
5000 μA 5000 μA ± 5.5 μA
[9]
± 3 μA
[9]
16 Press button to go to
µA mode
[1]
500 μA500 μA, 1 kHz ± 4.2 μA ± 3.7 μA
500 μA, 20 kHz ± 15.8 μA ± 3.95 μA
5000 μA 5000 μA, 1 kHz ± 42 μ 37.0 μA
5000 μA, 20 kHz ± 0.156 mA ± 39.5 μA
17 Turn the rotary switch to the
position
50 mA 50 mA ± 0.105 mA
[9]
± 80 μA
[9]
440 mA 400 mA ± 0.93 mA
[9]
± 0.71 mA
[9]
18 Press button to go to
mA mode
[1]
50 mA 50 mA, 1 kHz ± 0.42 mA ± 0.37 mA
50 mA, 20 kHz ± 1.56 mA ± 0.395 mA
440 mA 400 mA, 45 Hz ± 3.4 mA ± 3.0 mA
400 mA, 1 kHz ± 3.4 mA ± 3.0 mA
Caution: Connect the calibrator to handheld multimeter's A and COM terminals before applying 5A and 10A
5 A 5 A ± 16 mA ± 16 mA
10 A
[10]
10 A ± 40 mA ± 35 mA

Table of Contents

Related product manuals