EasyManua.ls Logo

Agilent Technologies U1252B - Product Characteristics

Agilent Technologies U1252B
191 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
152 Agilent U1251B/U1252B User’s and Service Guide
7 Specifications
Product Characteristics
POWER SUPPLY
Battery type:
9 V size Ni-MH rechargeable battery, 7.2 V nominal voltage
9 V size Ni-MH rechargeable battery, 8.4 V nominal voltage
9 V Alkaline battery (ANSI/NEDA 1604A or IEC 6LR61)
9 V Carbon-zinc battery (ANSI/NEDA 1604D or IEC6F22)
Battery life:
8 hours typical (based on a fully charged Ni-MH 300 mAH battery for DC
voltage measurement)
14 hours typical (based on a new 9 V Alkaline battery for DC voltage
measurement)
Charge time:
Less than 220 minutes in an environment of 10 °C to 30 °C. if the battery has
been deep-discharged, a prolonged charging time is required in order for the
battery to return to full capacity
POWER CONSUMPTION
105 mVA / 420 mVA (with backlight usage) maximum (U1251B)
165 mVA / 480 mVA (with backlight usage) maximum (U1252B)
DISPLAY
Both primary and secondary displays are 5-digit liquit crystal display (LCD)
with maximum reading of 50,000 counts
Automatic polarity indication
OPERATING ENVIRONMENT
Temperature: Full accuracy from –20 °C to 55 °C.
Humidity: Full accuracy up to 80% RH (relative humidity) for temperatures up
to 35°C, decreasing linearly to 50% RH at 55 °C.
Altitude:
0 to 2000 meters complying with IEC 61010-1 2nd Edition CAT III, 1000 V/
CAT IV, 600 V.
Pollution degree II
STORAGE COMPLIANCE
–40 °C to 70 °C with battery removed