Performance Tests and Calibration 6
Agilent U1251B/U1252B User’s and Service Guide 142
For an adjustment that is successful, the secondary display
will briefly show PASS. An adjustment failure is indicated
by a long beep and a calibration error number appears in
the secondary display. The primary display remains at the
current Cal Item. Check the input value, range, function,
and entered adjustment value to fix the problem. Repeat
the adjustment procedure.
9 Repeat steps 1 through 8 for each adjustment point.
10 Verify the adjustments using the “Performance Verification
Tests on page 126 and Table .