EasyManua.ls Logo

Agilent Technologies U1253B

Agilent Technologies U1253B
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests and Calibration 6
U1253B User’s and Service Guide 153
Changing Calibration Security Code
From front panel
1 After unsecuring the instrument, press for more
than 1 second to enter Calibration Security Code setting
mode.
2 The existing code will be shown on the secondary display,
for example, “CSC:C 1234”, where the character “C”
signifies “change”.
3 Press or to start and choose which digit to edit,
and press or to edit the value. (To exit without
changing the code, press for more than 1 second.)
4 Press (Save) to save the new security code.
5 If the new calibration security code has been successfully
stored, the upper left corner of the secondary display will
momentarily show the word “PASS”.
Please refer to Figure 6- 5 on page 154.

Table of Contents

Other manuals for Agilent Technologies U1253B

Related product manuals