166 U1253B User’s and Service Guide
6 Performance Tests and Calibration
Table 6-4 List of calibration items
Function Range Calibration item
[1]
Reference input
AC V
(with rotary switch at
and
[2]
)
5 V 0.3000 V (1 kHz)
3.0000 V (1 kHz)
3.0000 V (10 kHz)
0.3 V, 1 kHz
3 V, 1 kHz
3 V, 10 kHz
50 V 3.000 V (1 kHz)
30.000 V (1 kHz)
30.000 V (10 kHz)
3 V, 1 kHz
30 V, 1 kHz
30 V, 10 kHz
500 V 30.00 V (1 kHz)
300.00 V (1 kHz)
300.00 V (10 kHz)
30 V, 1 kHz
300 V, 1 kHz
300 V, 10 kHz
1000 V 30.0 V (1 kHz)
300.0 V (1 kHz)
300.0 V (10 kHz)
(done for this function; change
rotary switch position or press
to select the next function
that requires calibration)
30 V, 1 kHz
300 V, 1 kHz
300 V, 10 kHz
DC V Short SHORT Dual banana shorting plug with
copper wire
5 V 3.0000 V 3 V
50 V 30.000 V 30 V
500 V 300.00 V 300 V
1000 V 1000.0 V
(done)
1000 V
DC mV Short SHORT Dual banana shorting plug with
copper wire
50 mV 30.000 mV 30 mV
500 mV 300.00 mV 300 mV
1000 mV 1000.0 mV
(done)
1000 mV