182 U1253B User’s and Service Guide
7 Specifications
AC Specifications
Table 7-2 Accuracy specifications ± (% of reading + number of LSD) for true RMS AC voltage
Function Range
[5]
Resolution
Accuracy
[1][2][3]
20 Hz to 45
Hz
45 Hz to 1
kHz
1 kHz to 5
kHz
5 kHz to 15
kHz
15 kHz to
100 kHz
[4]
Voltage
50.000 mV 0.001 mV 1.5 + 20 0.4 + 40 0.7 + 40 0.75 + 40 3.5 + 120
500.00 mV 0.01 mV 1.5 + 60 0.4 + 25 0.4 + 25 0.75 + 40 3.5 + 120
1000.0 mV 0.1 mV 1.5 + 60 0.4 + 25 0.4 + 25 0.75 + 40 3.5 + 120
5.0000 V 0.0001 V 1.5 + 60 0.4 + 25 0.6 + 25 1.5 + 40 3.5 + 120
50.000 V 0.001 V 1.5 + 60 0.4 + 25 0.4 + 25 1.5 + 40 3.5 + 120
500.00 V 0.01 V 1.5 + 60 0.4 + 25 0.4 + 25 - -
1000.0 V 0.1 V 1.5 + 60 0.4 + 40 0.4 + 40 - -
Notes for AC voltage specifications:
1 Input impedance: Refer to Ta b l e 7 - 1 7 .
2 These specifications are defined for signal input > 5% of range.
3 Crest factor ≤ 3.0 at full scale, 5.0 at half scale except the 1000 mV and 1000 V ranges where it is 1.5 at full scale, 3.0 at
half scale. For non-sinusoidal waveform, add 0.1% of reading ± 0.3% of range.
4 Additional error to be added for frequency > 15 kHz and signal input < 10% of range: 3 counts of LSD per kHz.
5 2% over-range on all ranges except AC 1000 V range.