LS User Manual 35
9.0 SETTING UP A NEW TEST
9.1 PRE-DEFINED TEST SETUPS
The LS is supplied with three pre-dened test setups and one for new test setups, these are:
LIMIT_TEST A test which is carried out to a limit. The limits are dened in mm or inches,
% elongation (after setting gauge length), Newtons, Kg force, lbs force, Mega
Pascal’s Mpa (have entered the sample dimensions), Kg force per mm² and lb
force per inch²
CYCLE_TEST A cycle is from home to a limit (as set out above) and back to home. If when
the test starts the crosshead is not at the home position, it will move to the
home position rst before going to the limit position. The machine can be set
to perform up to 10000 cycles per test.
BREAK_TEST A Break Test will either, stop at a limit as described above or when a break is
detected. There are 2 types of break detector. A % break detector which
detects when the load has dropped to 50% of the maximum peak load
detected and sharp break detector. The sharp break detector operates when
there has been a sharp change in load or direction between one load reading
and the next.
NEW_TEST Is a user dened test rather than one of the default settings which are detailed
on the next page.