10.0 ERROR CONDITIONS
10.1 ERROR CONDITIONS
On the LS, if an error occurs, a message will be displayed on the console describing the error. This mes-
sage must be cleared by pressing the ENTER key before proceeding. Some errors are displayed when
the machine is rst powered on and indicate faults that were detected during the power up sequence.
If the LS machine is being used with a computer and NEXYGENPlus™ software, similar messages will be
displayed on the computer by NEXYGENPlus™.
The ERROR messages are as follows:
Data Packet Error
This error indicates that an invalid data packet has been received across the USB interface. All data
transferred via USB is checked for errors and resent if any error occurs, so this message should never be
displayed. If it is seen it is indicative of other problems occurring either within the machine or at the PC.
Motor Drive Fault
This indicates that the motor drive system has reported an error condition.
Upper limit switch
The adjustable upper limit switch has been hit. This can be cleared by adjusting the upper limit or driving
off it using the jog down keys.
Lower limit switch
The adjustable lower limit switch has been hit. This can be cleared by adjusting the lower limit, or driving
off it using the jog up keys.
No Loadcell Available
An attempt has been made to drive the machine without a load cell connected. Ensure the load cell is
properly connected and try again.
Splinter Shield Open
The splinter shield tted to the machine is open. A test cannot be started until the splinter shield is closed.
Cell Overload
The load applied to the system is in excess of 110% of the load cell rating. Use the jog keys to remove the
load.
Safety Overload
The load applied to the load cell is in excess of the safety load limit in the global settings. This limit only
applies when the crosshead is moving downwards during return or under control of the jog keys.
Frame Overload
The load applied to the system is in excess of 105% of the frame rating. Use the jog keys to remove the
load.
Grip Overload
The load applied to the system is in excess of the grip protection limit in the global settings. This limit only
applies during a test.
Print unavailable. Power down and retry
An attempt has been made to print when the USB port is in use with a PC running NEXYGENPlus™.
After using the machine with NEXYGENPlus™, it is necessary to power down before it is possible to
print.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek LS1 Series and is the answer not in the manual?

Ametek LS1 Series Specifications

General IconGeneral
BrandAmetek
ModelLS1 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals