3-27
(9) Enter RF/Micro Conv maintenance menu, and open its 6th page :
1) Enter Cal menu by pushing Shift + 0 keys. Open second page of Cal menu by pushing more.
2) Enter maintenance menu with F6 (Maintenance) key.
3) Enter RF/Micro Converter maintenance menu with F2 (Mainte RF/Micro conv) key. Open the 6th page of RF/
Micro converter maintenance menu (Press more key 5 times).
(10) Set IF Gain-1 and IF Gain-2 values to 0.
1) Press F4 (IF Gain-1) + 0 + enter keys and on display appears a writing IF Gain1 set to 0.
2) Press F5 (IF Gain-2) + 0 + enter keys and on display appears a writing IF Gain2 set to 0.
(11) Adjust IF Gain 1 and IF Gain 2 so that marker reading becomes -12 dBm ±0.5 dB.
If the level is lesser than this, the level can be raised by increasing the Number of IF Gain-2 (F5) from 0 to 255 in
single whole numbers.
Press F5 + number (0 to 255) + enter keys.
If the level is greater than this, the level can be lowered by increasing the Number of IF Gain-1 (F4) from 0 to 255
in single whole numbers.
Press F4 + number (0 to 255) + enter keys.
[ Band 1+ ]
(12) Set Manual Band to Band 1+ :
1) Enter Frequency menu by pushing Frequency key, and open its second page with more key.
2) Enter Internal mixer Band menu with F1 (Internal Mix) key.
3) Set manual band to Band 1+ by pushing F4 (Manual Band 1+) key.
(13) Set the spectrum analyzer to :
Center frequency : 7.25 GHz
Span : 200 MHz
Set the 6769B output frequency to 7.25 GHz (CW).
(14) Adjust the 6769B output level to make power meter reading -10 dBm at end of cable feeding the signal to the spectrum
analyzer, and then connect the cable to the spectrum analyzers RF Input.
(15) Tune the spectrum analyzers pre-selector (refer to procedure (7)).
(16) Enter RF/Micro Conv maintenance menu, and open its 6th page (refer to procedure (9)).
(17) Set IF Gain-1 and IF Gain-2 values to 0 (refer to procedure (10)).
(18) Adjust IF Gain 1 and IF Gain 2 so that marker reading becomes -12 dBm ±0.5 dB (refer to procedure (11)).