Section 3 MS2667C
3-28
[ Band 2+ ]
(19) Set Manual Band to Band 2+ :
1) Enter Frequency menu by pushing Frequency key, and open its second page with more key.
2) Enter Internal mixer Band menu with F1 (Internal Mix) key and open its second page with more key.
3) Set manual band to Band 2+ by pushing F2 (Manual Band 2+) key.
(20) Set the spectrum analyzer to :
Center frequency : 11.65 GHz
Span : 200 MHz
Set the 6769B output frequency to 11.65 GHz (CW).
(21) Adjust the 6769B output level to make power meter reading -10 dBm at end of cable feeding the signal to the spectrum
analyzer, and then connect the cable to the spectrum analyzers RF Input.
(22) Tune the spectrum analyzers pre-selector (refer to procedure (7)).
(23) Enter RF/Micro Conv maintenance menu, and open its 6th page (refer to procedure (9)).
(24) Set IF Gain-1 and IF Gain-2 values to 0 (refer to procedure (10)).
(25) Adjust IF Gain 1 and IF Gain 2 so that marker reading becomes -12 dBm ±0.5 dB (refer to procedure (11)).
[ Band 3+ ]
(26) Set Manual Band to Band 3+ :
1) Enter Frequency menu by pushing Frequency key, and open its second page with more key.
2) Enter Internal mixer Band menu with F1 (Internal Mix) key and open its second page with more key.
3) Set manual band to Band 3+ by pushing F3 (Manual Band 3+) key.
(27) Set the spectrum analyzer to :
Center frequency : 18.8 GHz
Span : 200 MHz
Set the 6769B output frequency to 18.8 GHz (CW).
(28) Adjust the 6769B output level to make power meter reading -10 dBm at end of cable feeding the signal to the spectrum
analyzer, and then connect the cable to the spectrum analyzers RF Input.
(29) Tune the spectrum analyzers pre-selector (refer to procedure (7)).
(30) Enter RF/Micro Conv maintenance menu, and open its 6th page (refer to procedure (9)).
(31) Set IF Gain-1 and IF Gain-2 values to 0 (refer to procedure (10)).
(32) Adjust IF Gain 1 and IF Gain 2 so that marker reading becomes -12 dBm ±0.5 dB (refer to procedure (11)).