EasyManua.ls Logo

Anritsu s331c - Save on Event Menu

Anritsu s331c
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
File Management 4-4 File Menu
Site Master User Guide PN: 10580-00252 Rev. J 4-11
Save On Event Menu
Key Sequence: File > Save On Event
In Spectrum Analyzer mode, this menu is used to auto save measurements
to internal memory after:
...Crossing Limit: Toggling this submenu key to On will save the
measurement to internal memory when the measurement has crossed a
defined limit line created with the Limit menu.
...Sweep Complete: Toggling this submenu key to On will save the
measurement to internal memory after the current sweep is complete. If
Save Then Stop is toggled Off, a measurement will be saved after every
sweep.
Save Then Stop: Set this key to On to stop the sweep after a
measurement is saved. With this key Off and Sweep Complete On a
measurement is saved after every sweep.
Clear All: Pressing this key will turn Off the three save on event keys:
Crossing Limit
Sweep Complete
Save Then Stop
Figure 4-11. Save On Menu
Back
Clear All
Save On...
...Crossing Limit
On Off
...Sweep Complete
On Off
Save Then Stop
On Off

Table of Contents

Related product manuals