Chapter 3 Install the Colocated Layout
Perform the RNase P Experiment
45
Applied Biosystems StepOne
™
and StepOnePlus
™
Real-Time PCR Systems
Installation, Networking, and Maintenance Guide
Notes
5. Review the standard curve:
a. Select the Standard Curve tab.
b. Click the upper-right corner of the Plate Layout to select all wells.
c. Confirm that the R2 value is greater than or equal to 0.990.
If the R2 value is less than 0.990, repeat the RNase P experiment using a different
RNase P plate. If the problem persists, contact Applied Biosystems as explained in
“How to Obtain Support” on page x.
6. Click Finish, click Ye s when prompted to save the experiment.