15.3 Configuration commands
# Instruction format Specification Return format
1 Measuring time in seconds
Write WMtime%d [CR] %d = 30…300
Default: 60
Mtime:%d[s];CRC:z[CR][LF]
Read RMtime[CR] -
2 Pause time in seconds
Write WHtime%d[CR] %d = 1…86400
Default: 10
Htime:%d[s];CRC:z[CR][LF]
Read RHtime[CR] -
3 Operating
mode
Write SStartMode%d[CR] %d = 0: time-controlled
measurement (default)
%d = 1: Digital I/O
%d = 2: Key / RS232
%d = 3: Automatic
StartMode:%d;CRC:z[CR][LF]
Read RStartMode[CR] -
4 Autoparts: Number of particles if operating mode= Automatic
Write WAutoParts%d[CR] %d = 200…5000000
Default: 200
AutoParts:%d[-];CRC:z[CR][LF]
Read RAutoParts[CR] -
5 Start and stop of a measurement in operating mode "Key"
Start Start[CR] - Measuring[CR][LF]
Stop Stop[CR] - See return format read command
“RVal”
6 Volume flow in ml/min
Write WFlow%d[CR] %d = 0...400
0 = Automatic (default)
1…400 = Fixed value
Flow:%d[ml/min];CRC:z[CR][LF]
Read RFlow[CR] -
7 Automatic measurement value output via RS232
Write SAutoT%d[CR] %d = 0: deactivated (default)
%d = 1: activated
AutoT:%d;CRC:z[CR][LF]
Read - -
8 Standard to be displayed
Write SStd%d[CR] %d = 0: ISO 4406:17 (default)
%d = 1: SAE AS4059E
%d = 2: NAS 1638
%d = 3: GOST 17216
Std:%d;CRC:z[CR][LF]
Read RCon[CR] - See answer: „RCon“
9 Alarm type
Write SAlarmD%d[CR] %d = 0: Standard alarm
%d = 1: Filter mode
Default: 0
AlarmD:%d;CRC:z[CR][LF]
Read RCon[CR] - See answer: "RCon"
10 Limit value alarm ISO/SAE 4µm (depending on set standard)
Write WAlarm4%c[CR] ISO: %c = 0…28
0 = Alarm deactivated
Default: 0
SAE: %c = 000…12
000 = Alarm deactivated
Default: 000
Alarm4:%c[-];CRC:z[CR][LF]
Read RAlarm4[CR] -