# Instruction format Specification Return format
11 Limit value alarm ISO/SAE 6µm (depending on set standard)
Write WAlarm6%c[CR] ISO: %c = 0…28
0 = Alarm deactivated
Default: 0
SAE: %c = 000…12
000 = Alarm deactivated
Default: 000
Alarm6:%c[-];CRC:z[CR][LF]
Read RAlarm6[CR] -
12 Limit value alarm ISO/SAE 14µm (depending on set standard)
Write WAlarm14%c[CR] ISO: %c = 0…28
0 = Alarm deactivated
Default: 0
SAE: %c = 000…12
000 = Alarm deactivated
Default: 000
Alarm14:%c[-];CRC:z[CR][LF]
Read RAlarm14[CR] -
13 Limit value alarm ISO/SAE 21µm (depending on set standard)
Write WAlarm21%c[CR] ISO: %c = 0…28
0 = Alarm deactivated
Default: 0
SAE: %c = 000…12
000 = Alarm deactivated
Default: 000
Alarm21:%c[-];CRC:z[CR][LF]
Read RAlarm21[CR] -
14 Limit value alarm NAS
Write WAlarmNAS%c[CR] %c = 00…12
00 = Alarm deactivated
Default: 00
AlarmNAS:%c[-];CRC:z[CR][LF]
Read RAlarmNAS[CR] -
15 Limit value alarm GOST
Write WAlarmGOST%c[CR] %c = 00…17
00 = Alarm deactivated
Default: 00
AlarmGOST:%c[-];CRC:z[CR][LF]
Read RAlarmGOST[CR] -
16 Limit value alarm temperature in °C
Write WAlarmT%d[CR] %c = 0…85
0 = Alarm deactivated
Default: 0
AlarmT:%d[°C];CRC:z[CR][LF]
Read RAlarmT[CR] -
17 Current output
Write SAO1%d[CR]] %d = 0: deactivated
%d = 1: ISO/SAE 4μm
%d = 2: ISO/SAE 6μm
%d = 3: ISO/SAE 14μm
%d = 4: ISO/SAE 21μm
%d = 5: ISO/SAE sequential
(default)
%d = 6: NAS
%d = 7: GOST
AO1:%d;CRC:z[CR][LF]
Read RCon[CR] - See answer: „RCon“