12
001-15698-EN rev.12
Display Symbols
Following is a description of all the symbols that may appear on the display.
Battery Charge Level
• Several symbols showing a battery at a different charge level. These symbols
indicate status regarding the batteries and/or the regular charging system
(see "Battery Information" and "" on page 20 for details).
ECS Indicator
• On units equipped with an Enhanced Charging System (ECS) option, this
symbol shows up when the batteries are charging.
RTC (Return to Charge)
• This symbol indicates that the RTC function was activated by the hand
control.
RTH (Return to Home)
• This symbol indicates that the RTH function was activated by the hand
control.
PPP PPP (Pre-Programmed Position)
• This symbol indicates that the PPP function was activated by the hand
control.
Over Duty Cycle
• This symbol flashes when the lift is used for a long period without allowing it
to cool down. This security feature disables the function (Up, Sit/Lay or Left/
Right) that is being used beyond recommended duty cycle to prevent
damages.
• When the overheat protection is engaged, the function, which exceeded the
duty cycle, is disabled while all other functions remain active.
• The red light will stay on during a cooling period. In addition, a single beep
will be heard.
General Malfunction (Warning symbol accompanied with a numeric character)
• This symbol flashes when the lift encounters a malfunction.
• The lift needs to be inspected / repaired by a qualified service technician.
• The red light will stay on and a single beep will be heard.
Maintenance Required (Wrench symbol accompanied with a letter character)
• This symbol is shown for 7 seconds (accompanied with 2 beeps) when the
lift is waking up from sleep mode and flashes each 5 seconds to draw user's
attention.
• Contact your Arjo service centre for maintenance.
Overweight
• This symbol flashes when a transfer is attempted with a load exceeding the
lift's capacity.
• The red light will stay on (accompanied with 4 rapid beeps) until the patient
is lowered.
• The function is disabled while all other functions remain active.