Care and Maintenance
Cleaning
Cleaning of Lift and Accessories
Pay special attention to areas listed below that are most likely to contain dirt and microbes.
NOTE: It is recommended to clean the lift and its accessories between each patient use.
NOTE: Assure thorough cleaning and disinfection of the Emergency Cord. Make sure lift is ON.
CAUTION: To ensure a better rolling surface for the trolley wheels, clean the inside of the rail every 4
months. To do so, insert a damp cloth in the opening and slide it from one end of the rail to the other.
CAUTION: Do not drench the product, as this could damage electrical components and cause internal
corrosion. If a hot air dryer is used to dry the lift, the temperature must not exceed 80°C (176 °F). Do not
use methanol-based or any other type of solvents that may damage the finish or plastic parts.
WARNING: To avoid eye and skin damage, always use protective glasses and gloves. If contact
occurs rinse with plenty of water. If eyes or skin becomes irritated,
contact a physician. Always read the instructions for use and
material safety data sheet of the disinfectant.
Joint between
Latch
Frame and Cover
Hand-held
Controller
Strap
Controller
edge
of Spreader Bar
Spreader Bar and Strap
Cord
Back and Handle
Emergency
1) Use a cloth soaked with water to remove
excessive residues from the lift and its
accessories from top to bottom.
2) Use a damp cloth with warm water mixed with
a cleaning fluid.
3) Scrub the equipment from top to bottom to
remove any deposits.
4) Use a clean damp cloth with clear water to
wipe off all spots of cleaning fluids.
5) Then wipe clean with a dry cloth.