Disinfection
Wipe the Maxi Sky 2 and accessories using a solution containing one of the compatible disinfectants below.
Disinfection methods should comply with the local or national guidelines (Decontamination of Medical
Devices), depending on the Healthcare Facility or country of use.
If uncertain, consult local Infection Control specialist.
NOTE: It is recommended to clean the lift and its accessories before disinfection.
NOTE: After thorough cleaning and/or disinfection of the Emergency Cord, assure the lift is still "ON".
Disinfection Agents for
Lift and Accessories
Disinfection Agents
Quaternary Ammonium ≤ 0.21% (2,100 ppm)
Isopropyl Alcohol ≤ 70%
Phenolic
≤ 1.56% (15,600 ppm)
Bleach ≤ 1% (10,000 ppm)
Peracetic Acid ≤ 0.25% (2,500 ppm)
Hydrogen Peroxide ≤ 6%
Basic Ceiling Lifter model (woven nylon strap & clevis pin Spreader Bar)
√ √ √
Maxi Sky 2 Ceiling Lifter model (woven nylon strap & Quick Connect Spreader Bar)
√ √ √
Infection Control model (coated strap & Quick Connect Spreader Bar)
√ √ √ √ √ √
Other Accessories for all three Maxi Sky 2 models
√ √ √ √ √ √
*
**
* The cover for this particular model may be vinyl padding or plastic. If vinyl, follow disinfection instructions for the Basic model.
** Handset attachment cable may become slightly grimy after repeated contact with bleach or hydrogen peroxide.