234
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Current Display (DC Only)
Range 1
0.0uA ~ 32.0uA, DC
Resolution
0.1uA
Base Accuracy
DC : ±(2% of reading + 3counts)
(>10uA)
Range 2
7.8uA ~ 130.0uA, DC
Resolution
0.1uA
Base Accuracy
DC : ±(2% of reading + 3counts)
(>10uA)
Range 3
31.5uA ~ 525.0uA, DC
Resolution
0.1uA
Base Accuracy
DC : ±(2% of reading + 3counts)
(>10uA)
Range 4
126.0uA ~ 999.9uA, 1000uA ~
2100uA, DC
Resolution
1uA
Base Accuracy
DC : ±(2% of reading + 3counts)
(>10uA)
Range 5
504uA ~ 999.9uA, 1000uA ~ 8400uA,
DC
Resolution
1uA
Base Accuracy
DC : ±(2% of reading + 3counts)
(>10uA)
Range 6
600uA ~ 8399uA, 8.40mA ~ 10.00mA,
DC
Resolution
1uA, 0.01mA
Base Accuracy
DC : ±5% of reading (0.01mA -
10.00mA )
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Current Display (DC Only)
Range 1
N/A
Resolution
N/A
Additional Error
N/A
Range 2
0.0uA ~ 7.8uA, DC

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Associated Research OMNIA II 8207 and is the answer not in the manual?

Associated Research OMNIA II 8207 Specifications

General IconGeneral
BrandAssociated Research
ModelOMNIA II 8207
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals