28
follows: Setup Tests, Perform Tests, Results Summary, and Results.
11. ENTER KEY: Key used to finalize parameter entries. The ENTER key may also
be used to scroll the highlighted area to different parameters in the parameter
setting screens.
12. SENSE + TERMINAL: Connector used to attach the + sense lead for 4-wire
Kelvin current measurement. This connection provides for accurate current
measurement during the Ground Bond test.
13. CURRENT OUTPUT TERMINAL: Connector used to attach the high current
output lead, adapter box high current lead or test fixture high current lead to the
instrument. This connection provides the output current for the ground bond and
continuity.
14. HIGH VOLTAGE INDICATOR: This indicator flashes to warn the operator that
high voltage is present at the high voltage output terminal.
15. RETURN OUTPUT TERMINAL: Connector used to attach the return test lead,
adapter box return lead or test fixture return lead to the instrument. This
connection provides the return current path.
16. SENSE - TERMINAL: Connector used to attach the - sense lead for 4-wire Kelvin
current measurement. This connection provides for accurate current
measurement during the Ground Bond test.
17. HIGH VOLTAGE OUTPUT TERMINAL: Connector used to attach the high
voltage test lead, adapter box high voltage lead or test fixture high voltage lead to
the instrument. This connection provides the high voltage used during a Hipot test.