Chapter 4 – Functions of the instrument
UM-23100B-U User manual ATEQ F CLASS Page 47/90
c) Ballistic
This fill type enables fluctuation in the air pressure (filling parts with a high level of
deformation) and in particular allows the maximum fill limit to be exceeded without the
cycle stopping and an error message being displayed. However, in order to move to
stabilisation mode, the test pressure must be correct.
2.1.3. 8) Blow mode
When this option is activated, the instrument fills the part to low pressure and this is
maintained between 2 cycles. This option is used when there is risk of increased damp
or dust when testing very dirty parts.
The blow pressure is set using a second regulator.
Note: this option is automatic and permanent, and is incompatible with the pre-fill type
function.
2.1.3. 9) Fill regulator
This function provides a choice of which of two regulators to use for the fill (1 or 2).