Wiring for Remote Configuration
CONNECT OR SEPARATE SIGNAL CONNECTION CABLE ONLY WHEN THE UNIT HAS BEEN SWITCHED OFF.
From
Detector
To Amplier
13 Shield
12 Coil
11 Coil
44 Shield
40 Empty Pipe
44 Shield
46 Electrode
44 Shield
45 Electrode
Junction Box
CS C2 C1
E1
ES
E2
RS
EP
ES
11 Coil
12 Coil
13 Shield
Shield (cut o)
Green (11)
Yellow (12)
Black (13) Overall Shield
45 Electrode
44 Shield
46 Electrode
44
40 Empty Pipe
44 Shield
White (45) Electrode
Black (44) Shield
Brown (46) Electrode
Pink (40)
Black (44) Shield
Gray (cut o)
Gray (cut o)
Black (44) Shield
Pink (40)
Brown (46) Electrode
Black (44) Shield
White (45) Electrode
Black (13) Overall Shield
Yellow (12)
Green (11)
Shield (cut o)
Figure 20: Wiring for remote configuration
Remote style M2000 transmitter models can be ordered with standard cables measuring 15, 30, 50,100 and 150 feet. In
addition, cables up to 500 feet are available.
From Junction Box
To M2000
Transmitter
Connection No. Description Wire Color Connection Label
11 Coil Green C1
12 Coil Yellow C2
13 Main Shield Black (Red Ferrule) CS
45 Electrode White E1
44* Electrode Shield Black ES
46 Electrode Brown E2
40 Empty Pipe Pink EP
44* Empty Pipe Shield Black ES
*Connections with the No. 44 are lying on the same potential.
Empty Pipe Detection Considerations
Take into account the following cable length and conductivity requirements if you are using empty pipe detection.
Cable Length (Feet) Minimum Conductivity Required (µS/cm)
0* 5
100 20
500 100
* Meter Mount
Wiring
Page 20 March 2022MAG-UM-01272-EN-21