USING THE MAIN MENU PROGRAMMING OPTIONS
The following programming options are available from the Main Menu:
Meter Setup
Measurements
Inputs/Outputs
Clear Totals
Communications
Advanced
Info/Help
Language Select
In the section that follows, the applicable security level for each menu option is indicated as follows:
A
S
U
Administrative Service User
Options that can be set at Quick Setup are indicated with:
The factory default values are shown, enclosed in brackets.
OTE: N Options labeled [Factory Set] should not be changed without specific directions from authorized
BadgerMeter personnel.
Meter Setup
scale factor
[0.0%]
S
Changing the scale factor lets you adjust the meter accuracy without disturbing factory-set
parameters. You can tune the meter to meet changing application requirements. For example, if the
meter is under registering by 0.5%, set the scale factor to +0.5%. If the meter is over registering by
0.5%, the scale factor to –0.5%.
To set the scale factor, follow these steps from the Meter Setup menu:
1. Select scale factor, to open the scale factor display.
2. Select a value for scale factor.
3. Press [E], to save the new value and return to the Meter Setup menu.
Empty Pipe Detect
[ON]
S
When set to On, Empty Pipe Detect indicates to the outputs and the display that the meter is not
completely filled. When set to Off, empty pipe detect is disabled.
Enabling empty pipe detect requires a one-time calibration. Calibration is described in the Advanced
menu section under Empty Pipe Cal.
To set Empty Pipe Detect, follow these steps from the Meter Setup menu:
1. Select Empty Pipe Detect to view the Empty Pipe Detect display.
2. Position the arrow next to On or Off.
3. Press [E] to save the setting and return to the Meter Setup menu.
Power Line Freq
[Region]
S
Power Line Freq provides measuring immunity to industrial noise from a power supply feed.
To set Power Line Frequency, follow these steps from the Meter Setup menu:
1. Select Power Line Freq to view the Power Line Frequency display.
2. Position the arrow next to 50 Hz or 60 Hz.
3. Press [E] to save the setting and return to the Meter Setup menu.
Using the Main Menu Programming Options