Meter Setup
Excitation Freq
[Factory Set]
A
Use Excitation Freq to configure the DC excitation of the coils. Supported frequencies are dependent on
the configured power line frequency:
50 Hz 60 Hz
1 Hz 1 Hz
3.125 Hz 3.75 Hz
6.25 Hz 7.5 Hz
12.5 Hz 15 Hz
To change Excitation Frequency, follow these steps from the Meter Setup menu:
1. Select Excitation Freq to view the Excitation Frequency display.
2. Position the arrow to select a frequency.
3. Press [E] to save the setting and return to the Meter Setup menu.
Pipe Diameter
[Factory Set]
A
If the transmitter is replaced, verify that the pipe diameter matches the installed pipe size.
To change Pipe Diameter, follow these steps from the Meter Setup menu:
1. Select Pipe Diameter to open the Pipe Diameter display.
2. Position the arrow to select a diameter.
3. Press [E] to save the setting and return to the Meter Setup menu.
Sensor Factor
[Factory Set]
A
Use Sensor Factor to compensate for accuracy error as a result of the installed sensor. If accuracy
adjustment of the meter is required, see the scale factor parameter.
If the transmitter is replaced, this parameter must be reprogrammed with the original sensorfactor.
Sensor Offset
[Factory Set]
A
Use Sensor Offset to compensate for accuracy error as a result of the installed sensor. If accuracy
adjustment of the meter is required, see the scale factor parameter.
OTE: N Changes of the sensor offset have an impact on meter accuracy at low flow.
Using the Main Menu Programming Options