Error Code Description Recommended Solution
10010 A connection to the sensor could
not be established. One or more
required channels failed to connect.
Verify the sensor is energized
Verify the sensor communication cable is connected
Verify all inline network equipment is energized, properly
configured and connected
Verify all required network equipment cables are connected
Verify the network adaptor is properly
configured
Verify the sensor IP address is valid
Verify there are no other devices using the same IP address
Verify the sensor subnet is valid
Verify the sensor can be successfully pinged
Verify the network
firewall settings support connection to the
sensor
10011 A connection to the sensor could
not be established. One or more
required channels failed to connect
Verify the sensor is energized
Verify the sensor communication cable is connected
Verify all inline network equipment is energized, properly
configured and connected
Verify all required network equipment cables are connected
Verify the network adaptor is properly
configured
Verify the sensor IP address is valid
Verify there are no other devices using the same IP address
Verify the sensor subnet is valid
Verify the sensor can be successfully pinged
Verify the network
firewall settings support connection to the
sensor
10012 A connection to the sensor could
not be established. A timeout
occurred connecting required
channels.
Verify the sensor is energized
Verify the sensor communication cable is connected
Verify all inline network equipment is energized, properly
configured and connected
Verify all required network equipment cables are connected
Verify the network adaptor is properly
configured
Verify the sensor IP address is valid
Verify there are no other devices using the same IP address
Verify the sensor subnet is valid
Verify the sensor can be successfully pinged
Verify the network
firewall settings support connection to the
sensor
10020 A connection to the sensor could
not be established. One or more
required channels failed to connect.
Verify the sensor is energized
Verify the sensor communication cable is connected
Verify all inline network equipment is energized, properly
configured and connected
Verify all required network equipment cables are connected
Verify the network adaptor is properly
configured
Verify the sensor IP address is valid
Verify there are no other devices using the same IP address
Verify the sensor subnet is valid
Verify the sensor can be successfully pinged
Verify the network
firewall settings support connection to the
sensor
10021 A connection to the sensor could
not be established. One or more
required channels failed to connect.
Verify the sensor is energized
Verify the sensor communication cable is connected
Verify all inline network equipment is energized, properly
configured and connected
Verify all required network equipment cables are connected
Verify the network adaptor is properly
configured
Verify the sensor IP address is valid
Verify there are no other devices using the same IP address
Verify the sensor subnet is valid
Verify the sensor can be successfully pinged
Verify the network
firewall settings support connection to the
sensor