PN A51387AB
1-19
Analyzer Overview
Preparing the Analyzer for Sample Runs
1
NOTE
For aperture tubes up to 100 µm, you can use an Accuvette® ST (20 mL). Use an ST Beaker for any
aperture tube size larger than 100 µm.
4. Add a few drops of the calibrator, control sample, or sample to the electrolyte solution.
• For more information on calibration, see Calibrating the Aperture, page 3-10.
• For more information on control sample runs, see Running a Concentration Control Sample,
page 3-33.
• For more information on running a sample analysis, see Analyzing a Sample, page 6-1.
5. If necessary, use the platform release and lower the platform to the loading position.
6. If necessary, turn the platform so that the correct side (ST Beaker or Accuvette® ST) is on top
(see Sample Platform, page 1-9).
NOTE
The sample platform has two usable sides. The ST Beaker side has a built-in guide to ensure that
a Multisizer 4 Smart Technology Beaker of any size is in the correct position when you raise the
platform. The Accuvette® ST side has a smaller guide for Accuvette STs. The guides ensure that
the ST Beaker or Accuvette® ST will be in the correct position relative to the aperture tube,
electrode, and stirrer when you raise the platform.
7. Place the ST Beaker or Accuvette® ST on the sample platform.
8. Raise the platform.
NOTE
It is not necessary to use the platform release to lift the platform into position.