EasyManua.ls Logo

Beckman Coulter Multisizer 4 - Step 10: Verify Calibration

Beckman Coulter Multisizer 4
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN A51387AB
3-31
Installing and Calibrating an Aperture Tube
The Change Aperture Tube Wizard
3
4. When the process is complete, the Status Panel displays READY. The noise level is displayed as
a percentage of aperture size below the Measure noise level step (in grey text). The Wizard
automatically highlights the next step, Verify Calibration (page 3-31).
Step 10: Verify Calibration
This step in the Change Aperture Tube Wizard assumes that you have already calibrated the
currently installed aperture tube (page 3-10). The Multisizer 4 software will automatically verify the
calibration of the aperture tube using the serial number you entered in Step 1: Select the new aperture
tube, page 3-23.
If you have not calibrated the aperture tube, exit the Wizard and calibrate the aperture (page 3-10).
To verify aperture tube calibration:
1. In the Change Aperture Tube Wizard, click
(
Next
)
until you reach the step Verify Calibration.
2. To the right of the Verify Calibration step, click
(
Verify
)
. The Calibrator Size window opens.
3. Enter the appropriate calibrator size.
4. Prepare the Analyzer for calibration.
5. In the Calibrator Size window, click
(
Start
)
. For more information on the verification
procedure, see Verifying the Calibration, page 3-18.

Table of Contents