EasyManua.ls Logo

Beckman Coulter Multisizer 4 - Starting the Analyzer; Preparing the Analyzer for Sample Runs

Beckman Coulter Multisizer 4
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN A51387AB
1-18
Analyzer Overview
Starting the Analyzer
Starting the Analyzer
To start the Analyzer:
1. Power ON the Analyzer.
2. Empty the Waste Jar, if necessary.
3. Fill the Electrolyte Jar with electrolyte, if necessary.
4. Open the Multisizer 4 software and connect the software to the Analyzer.
5. On the Multisizer 4 software Main Menu bar, select
Run > Instrument Start-Up.
6. In the Instrument Start-Up dialog box:
a. Select checkboxes for each function you would like the Analyzer to perform on startup.
b. Click
(
Start
)
.
NOTE
The Multisizer 4 performs best when it has reached a steady working temperature. This
typically takes 15 minutes from powering on the Analyzer.
Preparing the Analyzer for Sample Runs
To prepare the analyzer for a calibration, control, or sample analysis run:
1. Start the Analyzer and connect the software to the Analyzer (see Starting the Analyzer,
page 1-18).
2. Wash the ST Aperture Tube and ST Beaker or Accuvette® ST (see Cleaning the Aperture Tube,
Beakers, and Accuvette® STs, page 1-15).
3. Fill an ST Beaker or Accuvette® ST with clean electrolyte solution.

Table of Contents