EasyManua.ls Logo

Beckman Coulter Multisizer 4 - Page 76

Beckman Coulter Multisizer 4
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN A51387AB
3-22
Installing and Calibrating an Aperture Tube
The Change Aperture Tube Wizard
To use the Change Aperture Tube Wizard:
Select
Run > Change Aperture Tube Wizard on the Main Menu bar. The Change Aperture Tube Wizard
window opens.
The following sections describe how to use the Change Aperture Tube Wizard:
• Step 1: Select the new aperture tube, page 3-23
• Step 2: Remove the Currently Installed Aperture Tube, page 3-26
• Step 3: Install the New Aperture Tube, page 3-26
• Step 4: Place a Beaker of Clean Electrolyte on the Platform, page 3-27
• Step 5: Close the Door, page 3-27
• Step 6: Fill the System, page 3-28
• Step 7: Adjust the Metering Pump, page 3-28
• Step 8: Set Current and Gain, page 3-28
• Step 9: Measure Noise Level, page 3-30
• Step 10: Verify Calibration, page 3-31
• Step 11: All Done, page 3-33

Table of Contents