Berthold Duo Xpert - Analogue Output Mapping (AO); Analogue output: AO Mapping

Berthold Duo Xpert
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Main Menu Device Setup LB 470 Level
136
56925BA2 Rev. 03, 11/2017
Fig.
118 Submenu "Analog Output"
Analogue output: AO Mapping
Device settings | Setup | Outputs | Analogue Output (AO) | AO Mapping
A function can be assigned to an analogue output in the submenu AO Mapping.
The current output signal is between 4 mA and 20 mA. The corresponding values
(e.g. level) can be freely assigned.
1
2
3
4
5
6
Selection arrow
Assignment
Input field lower value [%]
Input field upper value [%]
Display field Live value assigned function
Display field Live value current [mA]
Selection Assignment
Fig.
119 Analogue Output Mapping (Assignment)
The value which is assigned to a current output signal of 4 mA must be smaller
than the value which is assigned to that of 20 mA.
Process value
The values of the process value can be assigned in the
menu Signal Condition | PV Range or in the calibration
settings.
Raw count rate
Enter a count rate range that outputs the current
count rate at the current output.