7 Main Menu Device Setup LB 470 Level
56925BA2 Rev. 03, 11/2017
118 Submenu "Analog Output"
Analogue output: AO Mapping
Device settings | Setup | Outputs | Analogue Output (AO) | AO Mapping
A function can be assigned to an analogue output in the submenu ″AO Mapping″.
The current output signal is between 4 mA and 20 mA. The corresponding values
(e.g. level) can be freely assigned.
″
″
Input field lower value [%]
Input field upper value [%]
Display field Live value assigned function
Display field Live value current [mA]
Selection Assignment
119 Analogue Output Mapping (Assignment)
The value which is assigned to a current output signal of 4 mA must be smaller
than the value which is assigned to that of 20 mA.
The values of the process value can be assigned in the
menu Signal Condition | PV Range or in the calibration
settings.
Enter a count rate range that outputs the current
count rate at the current output.